Open original FDA record ↗Recall record
- FDA identifier
- Z-0123-2013
- Company in FDA record
- Beckman Coulter Inc.
- FDA classification date
- Oct 24, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Beckman Coulter has confirmed that the Shuttle to Barcode Reader and Cap Piercer alignment procedure cannot be performed on certain DxC configurations with v5.0.11 software installed. If the Shuttle to Barcode Reader becomes misaligned, the Operator will not be able to perform this alignment procedure on the DxC systems listed above. Software v5.0 .11 does not allow access to the Shuttle to Bar
Source facts
- Action
- Beckman Coulter sent an Urgent Field Safety Notice dated August 30, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that their Beckman Coulter Service Representative would contac them to reinstall the software. Customers were asked to complete and return the enclosed Response Form. For questions customers were instructed to call their Customers Support Center on the Beckman Coulter website at www.BeckmanCoulter.com/CustomerSupport/SupportRequest. For questions regarding this recall call 714-993-5321.
- Initiated
- 2012-09-03
- Root cause
- Software change control
- Terminated
- 2014-02-14
- Distribution
- Worldwide Distribution - USA including TX and OH. Interntionally to China and Turkey
- Recall number
- Z-0123-2013
- Recall status
- Terminated
- CDRH record ID
- 112676
- FDA classified
- 2012-10-24
- Recall event number
- 63143