Z-0123-2024: 8713051U, Infusomat Large Volume Pump, Wireless; Volumetric Infusion Pump System

Recall record Z-0123-2024 for B. Braun Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0123-2024
Company in FDA record
B. Braun Medical, Inc.
FDA classification date
Oct 25, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Upstream occlusion alarm may sound when no occlusion exists, and the device will stop pumping. Interruption to the infusion of high-risk medications may lead to hemodynamic instability requiring medical intervention to prevent permanent impairment to body structures or body functions. In some cases, this may be life-threatening or may lead to death.

Source facts

Action
On September 21, 2023, the firm began notifying customers of the issue via "URGENT MEDICAL DEVICE CORRECTION" letters. BBMI will schedule service to update impacted devices. Customers were provided with stickers to identify affected units while waiting for the service to be completed. In the meantime, customers were informed of measures that may aid in continued use of the pumps, such as only using non-impacted devices for the delivery of high-risk medications and placing the impacted units in care areas that do not administer high risk medications. Customers may contact BBMI by calling 1-800-627-7867 (1-800-627-PUMP) for technical support for potential alternate interim measures. ***Updated 8/7/24*** On August 7, 2024, firm identified additional units which were serviced using potentially impacted sensors. Serial numbers were added to the two models already in scope (Article Numbers 8713051U and 8713052U). Serial numbers from an additional model (Article No. 8713050U) also added and tracked under RES 95093.
Initiated
2023-09-21
Root cause
Component design/selection
Distribution
Domestic distribution nationwide. Foreign distribution to Canada.
Recall number
Z-0123-2024
Recall status
Open, Classified
CDRH record ID
203160
FDA classified
2023-10-25
Recall event number
92978

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗