Philips Medical Systems North America Co. Phillips
FDA classification date
Oct 22, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
Due to a potential geometry software lock up during a Digital Rotational Angiography (DRA) run, the stand and table movement may no longer be possible.
Source facts
Action
Philips Medical Systems North America Co. sent an URGENT - Device Correction, dated September 15, 2010, to their consignees. The letter identified the product, the problem, and the action to be taken. The consignees were informed that there was a potential geometry software lock up during a Digital Rotational Angiography (DRA) run, the stand and table movement may no longer be possible.
The consignees were advised that as geometry software errors happen intermittently, no actions can be performed by the consignees to minimize the risk except for not performing DRA procedures until the firm upgrades the software.
For question consignees were to contact the Philips Call Center at 800-722-9377 #5, #4, #1 and reference FCO 72200111 or FCO 72200125.
Initiated
2008-08-21
Root cause
Software design
Terminated
2010-11-19
Distribution
Devices were distributed to hospitals and medical facilities through out the U.S.
Recall number
Z-0124-2011
Recall status
Terminated
CDRH record ID
73599
FDA classified
2010-10-22
Recall event number
49488