Open original FDA record ↗Recall record
- FDA identifier
- Z-0124-2013
- Company in FDA record
- Heartsine Technologies, Limited
- FDA classification date
- Oct 27, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class I
Reason stated in the source
Certain Samaritan 300/300P PAD devices may experience one or both of the following conditions that could affect the ability to deliver therapy to a patient in an SCA event if needed: Issue 1 (On/Off Issue): the device may turn itself on and off without input from the user. When this occures, the normal sequence of audible instruction prompts that occurs when the device is turned on will be emit
Source facts
- Action
- HeartSine sent an Urgent: Medical Device Recall dated September 2012 to all affected customers. The letter identifed the affected product, problem and actions to be taken. A response form was included with the letter to be completed by customers and returned to the recalling firm's representative. In addition, HeartSine is providing a software upgrade to bring all users up to a more recent version of the software. For questions call 1-877-877-0147 or your service representative.The company issued an additional press release on September 13, 2013 in a further attempt to reach out to end customers that had not yet responded to the companys efforts.
- Initiated
- 2012-09-13
- Root cause
- Device Design
- Terminated
- 2015-07-10
- Distribution
- Worldwide Distribution and US (nationwide) in the following states: AK, AL, AP, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VA, WA, and WI.
- Recall number
- Z-0124-2013
- Recall status
- Terminated
- CDRH record ID
- 112662
- FDA classified
- 2012-10-27
- Recall event number
- 63139