Open original FDA record ↗Recall record
- FDA identifier
- Z-0124-2015
- Company in FDA record
- Cook Medical Incorporated
- FDA classification date
- Oct 24, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
MEDINVENTS, the manufacturer of the Spirotome" Soft-Tissue Biopsy Needle Set distributed by Cook Medical, has initiated a voluntary recall of these products. The recall has been initiated due to customer complaints that the protective caps were dislodged in the pouch. Further investigation disclosed potential dislodging of protective caps on the needle tips on other distributed products.
Source facts
- Action
- On 9/26/2014, URGENT: MEDICAL DEVICE RECALL notifications were sent to the affected distributors and institutions Risk Managers/Recall Administration with instructions for returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For questions, call Customer Relations at 1-800-457-4500 or 1-812-339-2235.
- Initiated
- 2014-09-26
- Root cause
- Package design/selection
- Terminated
- 2016-02-02
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of Arizona, California, Colorado, Hawaii, Iowa, Illinois, Massachusetts, Maryland, Michigan, Missouri, New Mexico, Pennsylvania, Texas, Virginia, Vermont, and Wisconsin, and the countries of AT; BE; CH; CY; CZ; DE; DK; ES; FI; FR; GB; HU; IC; IE; IS; IT; LB; LT; LU; NL; NO; PT; QA; SE; TR; and ZA.
- Recall number
- Z-0124-2015
- Recall status
- Terminated
- CDRH record ID
- 130489
- FDA classified
- 2014-10-24
- Recall event number
- 69417