Open original FDA record ↗Recall record
- FDA identifier
- Z-0124-2021
- Company in FDA record
- Quidel Corporation
- FDA record creation date
- Oct 19, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Instructions for use in the package insert for SARS Antigen FIA test were updated to remove all references to a specific viral transport medium product as well as the use of other viral transport media. Use of the specific viral transport medium products may result in False-positive results. Directly testing from specimens (NS or NP swab) is recommended.
Source facts
- Action
- On 09/17/20, "FIELD CORRECTIVE ACTION CORRECTION" notices dated 09/16/20 were emailed to customers. The firm called customers over the phone to convey the correction information to some customers. Customers were asked to take the following actions: 1) Discontinue the use of affected package inserts. 2) Access the updated instructions for use labeling which can be located by going to quidel.com/docs and selecting the Sofia SARS Antigen FIA Package Insert document. 3) In the event that your facility has used Remel viral transport media products (e.g., M4, M4RT, M6), please contact Quidel Technical Support to discuss your results. 4) Confirm receipt of this notice by returning a signed copy to customernotifications@quidel.com Please feel free to contact Technical Support at customernotifications@quidel.com or by calling 800.874.1517 (in the U.S.) with any questions or concerns that you may have regarding this issue.
- Initiated
- 2020-09-17
- Root cause
- Device Design
- Terminated
- 2022-08-09
- Distribution
- U.S. Nationwide Distribution: NY, KS, TX, AL, FL, MN, CA, IL, MA, MS, PA, GA, NH, MO, VA, MI, OR, WV, OK, NV, CT, TN, MD, LA, AR, HI, NJ, KY, VT, IA, RI, NM, OH, IN, AK, SC, NC, AZ, SD, NE, WI, and.CO.
- Recall number
- Z-0124-2021
- Recall status
- Terminated
- CDRH record ID
- 183267
- FDA record created
- 2020-10-19
- Recall event number
- 86276