Skip to content

Public FDA record

Z-0125-2022: Atellica IM 1600 Analyzer, SMN 11066000

Recall record Z-0125-2022 for Siemens Healthcare Diagnostics, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0125-2022
Company in FDA record
Siemens Healthcare Diagnostics, Inc
FDA record creation date
Oct 16, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Ancillary reagent packs used on Atellica IM 1300 and Atellica IM 1600 analyzers may overflow when pierced while on an analyzer located at an altitude of greater than 350m (1148ft) above sea level. As a result, customers at these altitudes may observe an increased number of Reagent Volume Check Errors.

Source facts

Action
On September 27, 2021, the firm initiated notifications to affected customers. Customer distribution included all customers located in a city at an altitude of 350 meters (1148 feet) or higher. Customers were informed that Ancillary reagent packs used on Atellica IM 1300 and Atellica IM 1600 analyzers may overflow when pierced while on an analyzer located at an altitude of greater than 350 (m) or 1148 (ft) above sea level. The letters provide directions for mitigating the potential for leakage. If you have any questions, contact your local Siemens Healthineers technical support representative.
Initiated
2021-09-28
Root cause
Under Investigation by firm
Distribution
Domestic distribution to AZ CA CO GA ID KS MO MT NC NE NM NV OH OK OR PA TN TX UT VA WA WI WY. Foreign distribution worldwide.
Recall number
Z-0125-2022
Recall status
Open, Classified
CDRH record ID
189543
FDA record created
2021-10-16
Recall event number
88792

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗