Spinal fixation system implant screws marked "DNI" are not intended to be implanted and may not have been manufactured in accordance with released manufacturing specifications, and if implanted the screws may not have adequate strength or performance.
Source facts
Action
On 8/21/24, recall notices were mailed and emailed to customers who were asked to do the following:
1) Examine affected devices and quarantine any screws with DNI markings.
2) Notify your staff about the recall. If affected devices were further distributed, provide the recall notice to these customers.
3) Complete and return the response forms via email to mdolan@xtantmedical.com
Initiated
2024-08-21
Root cause
Process change control
Distribution
US Nationwide distribution in the states of IL, WI, FL, PA, CA, TX, MI, MO, VA, OH, MA.
Recall number
Z-0125-2025
Recall status
Open, Classified
CDRH record ID
210129
FDA classified
2024-10-17
Recall event number
95395