Five lots of AMPLATZER Vascular Plugs were sterilized in a load that was not validated for the AMPLATZER Vascular Plug. However, only three of those five lots were distributed.
Source facts
Action
Urgent Recall letters, dated 09/01/05, were sent to affected customers via email, facsimile and with follow-up delivery via USPS. OUS customers were notified via Federal Express. The letter described the situation, the lots being recalled and included a response form that was to be faxed back to AGA Medical.
Initiated
2005-09-01
Root cause
Other
Terminated
2006-11-13
Distribution
OUS countries include: Australia, Germany, Italy, Korea, Malasia, Slovakia, Spain, Switzerland, Thailand, United Kingdom
Recall number
Z-0127-06
Recall status
Terminated
CDRH record ID
41698
FDA classified
2005-11-05
Recall event number
33323