In March, Siemens Healthcare Diagnostics issued an Urgent Medical Device Correction for decreased recovery of patient samples with the Dimension Vista® CSA & CSAE Flex® reagent cartridges when whole blood samples are run from Small Sample Containers (SSC). Siemens has confirmed that when whole blood samples are run from SSC, HbA1c values may exhibit a percent bias ranging from - 8.2% t
Source facts
Action
An Urgent Medical Device Correction, dated August 2014, was provided to all Dimension Vista HbA1c customers who received the affected lots to notify them of the issue, the potential risk to health, and actions to be taken by customers.
Initiated
2014-08-14
Root cause
Device Design
Terminated
2015-08-10
Distribution
Worldwide Distribution: US (nationwide) and country of: Canada.
Recall number
Z-0127-2015
Recall status
Terminated
CDRH record ID
130239
FDA classified
2014-10-24
Recall event number
69334