Z-0127-2017: Calix T PEEK Lumbar System, TLIF Trial

Recall record Z-0127-2017 for X Spine Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0127-2017
Company in FDA record
X Spine Systems Inc
FDA record creation date
Oct 18, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The trials and rasps used to prepare the surgical site for placement of the implant may become detached from the inserter assembly part and become lodged in the intervertebral space.

Source facts

Action
Consignees will be notified via email and courier and will be instructed to return the affected product. Upon receipt of the recalled trays, the trials and rasps will be removed from the tray and quarantined for evenutal destruction.
Initiated
2016-05-17
Root cause
Component design/selection
Terminated
2019-04-04
Distribution
Product was shipped to the following states: AZ, CA, CO, FL, GA, ID, IN, KY, LA, MD, MI, MO, NC, NV, NY, PA, PR, TX & WV. Product was also shipped to the following countries: Australia, England, Germany, Italy, Mexico, Portugal & United Kingdom.
Recall number
Z-0127-2017
Recall status
Terminated
CDRH record ID
147446
FDA record created
2016-10-18
Recall event number
74397

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗