Z-0128-2021: System, Nuclear Magnetic Resonance Imaging - Product Usage: is a whole body magnetic resonance scanner de

Recall record Z-0128-2021 for GE Healthcare, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0128-2021
Company in FDA record
GE Healthcare, LLC
FDA record creation date
Oct 20, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential electrical arc flash within the Power Gradient & RF (PGR) Cabinet.

Source facts

Action
On 09/08/20 the firm sent a Customer Letter to its consignees with the following instructions: You can continue to use the system. In the event the PGR cabinet needs to be serviced, power down the PGR at the Main Disconnect Panel (MDP) then follow MDP Lockout/Tagout from the current released service methods and documentation. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction
Initiated
2020-09-08
Root cause
Process change control
Terminated
2021-04-28
Distribution
Worldwide distribution - US Nationwide distribution including in the states of FL, KY, NC, NJ, NM, NY, TX, UT and WA. The countries of Canada, Hungary, India, Korea, China, Russia, Taiwan, Puerto Rico and Ubekistan.
Recall number
Z-0128-2021
Recall status
Terminated
CDRH record ID
183734
FDA record created
2020-10-20
Recall event number
86452

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗