Skip to content

Public FDA record

Z-0130-06: Baxter I-Pump Pain Management System infusion pump, product codes 2L3107, 2L3107K and 2L3107R, Baxter Hea

Recall record Z-0130-06 for Baxter Healthcare Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0130-06
Company in FDA record
Baxter Healthcare Corp.
FDA classification date
Nov 16, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Potential failures of the pump due to upstream occlusion alarms, electro-static discharge, system error 32 or 33, history printout errors and fluid volume in 'Use Previous Rx' when priming through the pump.

Source facts

Action
Baxter mailed Urgent Device Correction letters dated 2/10/05 to the domestic accounts on 2/10/05 via first class mail, to the attention of the Director of Nursing, advising them of the following hardware problems and software anomalies and providing workarounds for each. There are potential failures of the pump due to upstream occlusion alarms, electro-static discharge, Key-Only security mode (previously covered by letter dated 4/10/02), system error 32 or 33 codes, history printout errors and fluid volume in 'Use Previous Rx' when priming through the pump. The accounts were advised that as soon as revised software and hardware is available, Baxter would modify the pumps at no additional charge to the consignee. Any questions were directed to their Baxter Medication Delivery Product Information Center at 1-800-933-0303.
Initiated
2005-02-10
Root cause
Other
Terminated
2008-11-03
Distribution
Nationwide including Puerto Rico, and internationally through Baxter subsidiaries in Canada, Chile, Hong Kong, Kuwait and Saudi Arabia.
Recall number
Z-0130-06
Recall status
Terminated
CDRH record ID
37341
FDA classified
2005-11-16
Recall event number
31191

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗