- FDA identifier
- Z-0130-06
- Company in FDA record
- Baxter Healthcare Corp.
- FDA classification date
- Nov 16, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Potential failures of the pump due to upstream occlusion alarms, electro-static discharge, system error 32 or 33, history printout errors and fluid volume in 'Use Previous Rx' when priming through the pump.
Source facts
- Action
- Baxter mailed Urgent Device Correction letters dated 2/10/05 to the domestic accounts on 2/10/05 via first class mail, to the attention of the Director of Nursing, advising them of the following hardware problems and software anomalies and providing workarounds for each. There are potential failures of the pump due to upstream occlusion alarms, electro-static discharge, Key-Only security mode (previously covered by letter dated 4/10/02), system error 32 or 33 codes, history printout errors and fluid volume in 'Use Previous Rx' when priming through the pump. The accounts were advised that as soon as revised software and hardware is available, Baxter would modify the pumps at no additional charge to the consignee. Any questions were directed to their Baxter Medication Delivery Product Information Center at 1-800-933-0303.
- Initiated
- 2005-02-10
- Root cause
- Other
- Terminated
- 2008-11-03
- Distribution
- Nationwide including Puerto Rico, and internationally through Baxter subsidiaries in Canada, Chile, Hong Kong, Kuwait and Saudi Arabia.
- Recall number
- Z-0130-06
- Recall status
- Terminated
- CDRH record ID
- 37341
- FDA classified
- 2005-11-16
- Recall event number
- 31191