Reason stated in the source
Component missing: Some products were issued without thermal protection switches.
Source facts
- Action
- AMS issued Important: Medical Device Correction letters, dated April 17, 2007, to its US and international consignees by mail, instructing them to return an acknowledgement form. The letter informs customers that some HPS systems may lack a motor overload switch, that the firm believes that there is no immediate safety issue, and installation of the MOL switch will be conducted during normal preventive maintenance. AMS expects to complete installation of the switches by end of August 2007.
- Initiated
- 2007-04-17
- Root cause
- Process design
- Terminated
- 2010-03-23
- Distribution
- Worldwide Distribution - USA, Italy, France, Germany, Spain, UK, Denmark, Sweden, Australia, India, Japan, Greece, Switzerland, Saudi Arabia, Thailand, Turkey, Singapore, Taiwan/Hong Kong (China), and Austria.
- Recall number
- Z-0130-2008
- Recall status
- Terminated
- CDRH record ID
- 53454
- FDA classified
- 2008-03-20
- Recall event number
- 38336