Open original FDA record ↗Recall record
- FDA identifier
- Z-0131-2011
- Company in FDA record
- Ebi, Llc
- FDA classification date
- Oct 22, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
One of the two connector ends of the medium cross connector may not be fully secured to the rod after tightening. This may not be readily apparent to the operating surgeon, resulting in inadequate torsional stability.
Source facts
- Action
- Biomet Spine issued an Urgent Medical Device Recall letter dated September 16, 2010 to distributors and physicians, identifying the affected device and actions to be taken. Customers and distributors were requested to identify, remove, and immediately return the affected product to Biomet, following enclosed Product Return Instructions, and complete and return the Removal Tracking and Verification Form. Customers and distributors can contact Biomet at 1-800-526-2579.
- Initiated
- 2010-09-16
- Root cause
- Other
- Terminated
- 2012-06-01
- Distribution
- Worldwide Distribution: USA and the countries of Australia, Columbia, Great Britain, The Netherlands, New Zealand, and Switzerland.
- Recall number
- Z-0131-2011
- Recall status
- Terminated
- CDRH record ID
- 94512
- FDA classified
- 2010-10-22
- Recall event number
- 56791