Skip to content

Public FDA record

Z-0131-2023: LINEAR 7.5Fr. 40cc IAB with Accessories (APA), Model No. 0684-00-0480-02. intra-aortic balloon catheter.

Recall record Z-0131-2023 for Datascope Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0131-2023
Company in FDA record
Datascope Corporation
FDA classification date
Oct 24, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The balloon volume is incorrectly described as 34cc instead of 40cc on one of the three internal labels enclosed for hospital use. A labelling error may result in user confusion and/or delay in initiating therapy. Therapy delays may impact those patients who are more hemodynamically unstable.

Source facts

Action
On August 5, 2022, the firm notified affected customers via Urgent Field Safety Notice Medical Device Removal letters. Customers were instructed to examine their inventory for the affected product. Customers may return the product for full credit. The firm also informed customers that given current supply chain shortages that they are unable to commit to a timely replacement of affected catheters.
Initiated
2022-08-05
Root cause
Labeling Change Control
Distribution
No US distribution. Foreign distribution to Argentina, Brazil, Canada, Chile, Columbia, Ecuador, Egypt, France, Indonesia, Israel, Italy, Jordan, Kenya, Malaysia, Mexico, Netherlands, Pakistan, Poland, Portugal, South Korea, Russia, Spain, Turkmenistan, United Arab Emirates, United Kingdom, Uruguay.
Recall number
Z-0131-2023
Recall status
Open, Classified
CDRH record ID
196193
FDA classified
2022-10-24
Recall event number
90913

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.