Skip to content

Public FDA record

Z-0131-2025: Brand Name: NOVA ORTHO-MED INC. Product Name: MONARCH ROLLATOR RED Model/Catalog Number: 4329RD Product D

Recall record Z-0131-2025 for Nova Ortho-Med Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0131-2025
Company in FDA record
Nova Ortho-Med Inc
FDA classification date
Oct 17, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Firm identified a potential risk for a fall hazard. When the end user is readjusting from the slouched seated position to an upright seated position, there is the potential for the movement to cause the backrest to lift out of the backrest holder and disengage from the unit.

Source facts

Action
On 09/25/2024, the firm sent via email an "Urgent: Medical Device Correction" letter to inform customers that when the end user is readjusting from the slouched seated position to an upright seated position, there is the potential for the movement to cause the backrest to lift out of the backrest holder and disengage from the unit resulting in potential fall risks. Customers are instructed to: Immediately take the following steps to correct the affected units. 1. Print the NOVA MONARCH Model 4329 Product Recall Notice and post it in your store in a visible location to customers. 2. Check your inventory for 4329 model walkers (4329BL, 4329PL and 4329RD), isolate and hold the products until you contact NOVA ORTHO-MED INC for the next steps. 3. Contact all customers who purchased a 4329 model to notify them of the potential issue. Schedule a time that they can come in for repair or if the customer prefers NOVA can send the replacement components directly to their residence. 4. If you do not track serial numbers with customer purchase or do not have the customer information, please ensure that the NOVA MONARCH Model 4329 Product Recall Notice document is posted in a customer visible location. 5. Call NOVA customer service and request the replacement component to be sent to the required location. (Part# - HCT-432915) 6. Record the serial # s of the product that have been corrected and email them to NOVA Customer Service (info@novajoy.com) so that we can have for our records. If you prefer to call, then dial 1-800-557-6682. For questions or assistance - Contact Information: Monday through Friday, 8:30 AM to 5:00 PM, Pacific time Time. "Toll free number: 1-800-557-6682 "Email: info@novajoy.com
Initiated
2024-08-26
Root cause
Device Design
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL AZ, CA, CO, CT, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, WA, WI and WY. The country of Canada.
Recall number
Z-0131-2025
Recall status
Open, Classified
CDRH record ID
210090
FDA classified
2024-10-17
Recall event number
95375

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗