Z-0131-2026: Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (

Recall record Z-0131-2026 for Contego Medical LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0131-2026
Company in FDA record
Contego Medical LLC
FDA classification date
Oct 10, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter that may not open or may not fully open during sheath retraction and operation of the filter knob. The stent is used with a primary distal embolic protection device.

Source facts

Action
On 9/4/2025, communication and training to the field representatives started with in person letter distribution following immediately after. Customers were asked to do the following: Action Required 1) Inventory Check: Medtronic and Contego sales representatives will visit your facility to conduct an inventory count of devices from the affected lots. 2) Quarantine Affected Devices: Any identified devices will be removed from use and quarantined by the Medtronic or Contego team. 3) Return: Medtronic and Contego will remove affected devices. If replacements are desired, they can be provided during the same visit. 4) Confirmation: Sign and return the enclosed acknowledgment form to confirm receipt of this advisory and the completion of actions performed by Medtronic or Contego. 5) Please email the completed form to Contego Medical recall@contegomedical.com as soon as possible and retain a copy of the form to provide to the representative at the time of their visit. If you have questions, contact the firm's Recall Support Team at recall@contegomedical.com or call 919-459-7250 Extension 4.
Initiated
2025-09-04
Root cause
Device Design
Distribution
US Nationwide distribution in the states of FL, CO, CA, IL, NC, KY, TX, PA, DE, NJ, IN, NY, OH, LA, OK, SC, CT, WI, MA, KS, UT, MD, AL.
Recall number
Z-0131-2026
Recall status
Open, Classified
CDRH record ID
215842
FDA classified
2025-10-10
Recall event number
97522

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗