Under certain conditions, the product's wash buffer containing sodium azide can come in contact with the wash zone mechanism ground cables on either of the two wash zones. The copper in the ground strap may corrode and form an unstable chemical substance (metal azide) that may be sensitive to direct pressure and impact.
Source facts
Action
A product correction notice was issued to consignees on July 17, 2009. Field representatives were instructed to visit consignees, inspect cables and replace them
Initiated
2009-07-17
Root cause
Other
Terminated
2010-01-26
Distribution
Worldwide Distribution
Recall number
Z-0132-2010
Recall status
Terminated
CDRH record ID
84105
FDA classified
2009-11-09
Recall event number
52833
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.