Z-0132-2021: Yuno Mobile Operating Table - Product Usage: Intended for support and position the patient immediately be

Recall record Z-0132-2021 for Maquet Cardiovascular Us Sales, Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0132-2021
Company in FDA record
Maquet Cardiovascular Us Sales, Llc
FDA record creation date
Oct 20, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Plastic socket of the IEC 60320-1 socket pulled out of the metal cage and electronic parts exposed; where the socket was pulled out and main cable is still connected to the main socket (power source), there is a risk that a user or third person may suffer an electric shock when coming in contact with the exposed electronic components

Source facts

Action
Getinge issued Urgent Medical Device Recall on September 24, 2020, via FedEx 2 day delivery to US consignees. The letter states reason for recall, health risk and action to take: Per the IFU, section 9.1 (Yuno) and 8.1 (Yuno II), a visual and functional check is required to be performed by a trained person before operation. Follow the instructions in section 4.2.3.1of the Yuno and Yuno II IFU regarding connecting and disconnecting from the mains socket (power source). If the plastic socket of the cold-device plug is pulled out, do not try to put it back into the housing and if there are any exposed electronic components, do not touch them. Disconnect/detach the cable from the main socket, call your in-house technical service and arrange the repair of the table. Per IFU section 9.4 (Yuno) and 8.5 (Yuno II), a defective product may not be used and may not be repaired by yourself. A Getinge Service representative will contact you to schedule the correction. This work will be done at no cost to your facility Please complete and sign the attached URGENT MEDICAL DEVICE RECALL CORRECTION RESPONSE FORM (page 4) to acknowledge that you have received this notification. Return the completed form to Maquet/Getinge by e-mailing a scanned copy to MEERA2019@getinge.com or by faxing the form to 1-973-807-9205
Initiated
2020-09-24
Root cause
Under Investigation by firm
Terminated
2023-02-28
Distribution
Worldwide distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, BAHRAIN, BOLIVIA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, FINLAND, FRANCE GERMANY, HONG KONG, HUNGARY, INDIA, INDONESIA, IRAQ, ISRAEL, ITALY, JAPAN, KENYA, KOREA, KUWAIT, LATVIA, LEBANON, MACAU, MACEDONIA, MALAYSIA, MEXICO, NEW ZEALAND, OMAN, POLAND, QATAR, ROMANIA, RUSSIAN FEDERATION, RWANDA, SAUDI ARABIA, SERBIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TUNESIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM.
Recall number
Z-0132-2021
Recall status
Terminated
CDRH record ID
183828
FDA record created
2020-10-20
Recall event number
86495

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗