Skip to content

Public FDA record

Z-0132-2026: The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry fi

Recall record Z-0132-2026 for Insulet Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0132-2026
Company in FDA record
Insulet Corporation
FDA classification date
Oct 10, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.

Source facts

Action
Insulet notified both healthcare professionals and users on about 09/11/2025 via URGENT: Medical Device Correction letter. Instructions included that an update to the application will be pushed to all devices and should be implemented by users. Users were also requested to confirm receipt and acknowledgement of the notification but selecting the "acknowledgement button" in the email.
Initiated
2025-09-11
Root cause
Software design
Distribution
US Nationwide distribution.
Recall number
Z-0132-2026
Recall status
Open, Classified
CDRH record ID
216042
FDA classified
2025-10-10
Recall event number
97597

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗