The pin may fail because the metal used to make the pin does not meet the firm's hardness specification.
Source facts
Action
All affected U.S. accounts were sent a letter dated October 15, 2003 by certified mail, return receipt requested. The seven foreign initial consignee accounts were notified via electronic mail. Customers were asked to return non-implanted product and to advise the implanting surgeons of the problem so that they may take appropriate action in regard to their patients.
Initiated
2003-10-15
Root cause
Other
Terminated
2004-05-20
Distribution
The majority of the pins were distributed within the United States. Distribution also occured in seven foreign countries; Taiwan, Canada, Japan, Korea, Belgium, Australia and Finland.
Recall number
Z-0133-04
Recall status
Terminated
CDRH record ID
30115
FDA classified
2004-07-20
Recall event number
27484