Z-0133-2023: da Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon Console (SSC), Pa

Recall record Z-0133-2023 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0133-2023
Company in FDA record
Intuitive Surgical, Inc.
FDA classification date
Oct 25, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3) Yellow pedal associated with force bipolar is pressed, released with head in surgeon console 4) Blue pedal associated with bipolar is pressed, resulting in force bipolar energy delivery

Source facts

Action
On 09/20/22, correction notices were mailed to customers. Customers were advised of the following: 1) The system software version can be located on the Vision Cart Touchscreen in the Troubleshooting Tab. 2) Ensure all device users read and understand this notice. 3) Notice should be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. 4) Ensure proper instrument connections to the generator(s) and continue to follow all existing warnings and cautions found in user manuals 551400-15, USER MANUAL,SYS,P10,US,IS4000, 553873-07, USER MANUAL,I-A,MDR P10,IS4000-IS4200 and 554070-04, ADD,I-A,FORCE BIPOLAR,IS4000,IS4200. 5) Complete and return the acknowledgement form. 6) The recalling firm will follow-up with customers and schedule a software upgrade. 7) After the upgrade, inform all personnel that the software update/upgrade was completed. 8) If you have experienced this issue or require additional assistance, contact Technical Support: 800-876-1310, Option 1 (24/7) or customerservice@intusurg.com
Initiated
2022-09-20
Root cause
Software Design Change
Distribution
US: MS, PA, WI, KY, TN, MI, FL, GA, NC, NV, LA, OH, IL, MO, IN, CA, AK, MD, TX, IA, MA, NE, AZ, WA, CO, AL, SC
Recall number
Z-0133-2023
Recall status
Open, Classified
CDRH record ID
195973
FDA classified
2022-10-25
Recall event number
90826

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗