Incorrect dosage possible: A software anomaly in the CADD-Sentry Pro Administrator software (version B) where the values of a weight based protocol go to zero or round up to the next higher hundredths place, even if Administrator displayed the values correctly at data entry.
Source facts
Action
A Smiths Medical MD employee visited the consignees between August 14 and 31, 2007, to verbally notify them of the recall, collect the recall CD-ROM containing the affected software and replace with the updated software.
Initiated
2007-08-14
Root cause
Software design
Terminated
2008-10-31
Distribution
UT, MD, CO
Recall number
Z-0134-2008
Recall status
Terminated
CDRH record ID
64397
FDA classified
2008-02-26
Recall event number
44799