The Flo-Gard infusion Pump was released to the customer with OOS values for the Air Sensor Calibrations with the Air in Tubing test.
Source facts
Action
A Baxter Healthcare technician initially notified the sole customer on August 26, 2011 of the defective product and requested that the product be quarantined. On September 14, 2011 a technician visited the site to correct the problem observed on the Flo-Gard Infusion Pump. All required tests were performed and the results were found acceptable.
Initiated
2011-08-26
Root cause
Employee error
Terminated
2011-10-31
Distribution
Distributed only in Puerto Rico.
Recall number
Z-0134-2012
Recall status
Terminated
CDRH record ID
104641
FDA classified
2011-10-28
Recall event number
60168