This product was produced using a finishing process not identified as part of the manufacturing specification. The process used with the lots subject to this Recall was a bead blast process. (Bead Blasting vs. Shot Peened).
Source facts
Action
An Urgent Notice: Medical Device Recall, dated September 4, 2015, was sent to end users to alert them about the issue and possible risk to patients. Customers were requested to follow the actions to be taken for if they have affected product or not; complete the response form, and return affected product. Customers can call 610-719-5450 or a local Synthes Sales Consultant with any questions.
Initiated
2015-09-04
Root cause
Process control
Terminated
2016-08-26
Distribution
Nationwide Distribution
Recall number
Z-0134-2016
Recall status
Terminated
CDRH record ID
139928
FDA classified
2015-10-17
Recall event number
72134