Z-0134-2021: Siemens Dimension Vista Akaline Phosphatase (ALPI) Flex reagent cartridge - In vitro diagnostic test for

Recall record Z-0134-2021 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0134-2021
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA classification date
Oct 20, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Low outlier resultsgenerated with Alkaline Phosphatase (ALPI) Flex reagent cartridge.The maximum difference observed with a patient sample was approximately -27% at a concentration of 129 U/L (2.15 ukat/L). The low outlier results are generated from the last five (5) tests of the well

Source facts

Action
Siemens issued Urgent Medical Device Recall (UMDC # VC-20-02.A.US via Fed'X) and Urgent Field Safety Notice (UFSN # VC-20-02.A.OUS) on August 25, 2020 to all Dimension Vista customers who received the Dimension Vista ALPI flex reagent lots 19247AB, 19282BB and 19330BD to notify them of the issue, health risk and actions to take. Customers were instructed to: Discontinue use of and discard the Dimension Vista ALPI lots 19247AB, 19282BB, 19330BD. Review their inventory of these lots to determine their laboratorys replacement needs. Complete and return the attached form to this letter to request your no-charge replacement product(s). Complete and return the Field Correction Effectiveness Check Form attached to the letter within 30 days. If they received any complaints of illness or adverse events associated with the product listed in the letter, to immediately contact their local Siemens Remote Services Center or their local Siemens Technical Support Representative. Expanded Recall 3/12/21: Urgent Medical Device Correction (UMDC # VC-20-02.B.US) and Urgent Field Safety Notice (UFSN # VC-20-02.B.OUS) issued on March 12, 2021 to all Dimension Vista customers who received the Dimension Vista ALPI flex reagent lots 20337AB, 20311BA, 20275AA, 20218AA, 20203BA, 20170AA, 20125BA and 20078AB to notify them of the issue, and actions to be taken.
Initiated
2020-08-25
Root cause
Under Investigation by firm
Terminated
2023-05-04
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Australia, Austria, Belgium, Bahamas, Czech Republic, Denmark, France, Germany, Italy, Japan, Korea, Netherlands, Norway, New Zealand, Portugal, Slovakia, Slovenia, Spain, Switzerland.
Recall number
Z-0134-2021
Recall status
Terminated
CDRH record ID
183697
FDA classified
2020-10-20
Recall event number
86430

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗