Skip to content

Public FDA record

Z-0134-2023: ARIES SARS-CoV-2 Assay, REF: 50-10047

Recall record Z-0134-2023 for Luminex Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0134-2023
Company in FDA record
Luminex Corporation
FDA classification date
Oct 26, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A part defect with the wash 1 syringe of an SARS-CoV-2 assay could lead to incomplete extraction thereby producing an invalid or false negative assay result.

Source facts

Action
On 09/14/22, recall notices were emailed to customers who were asked to do the following: 1. Cease all use, and isolate or quarantine affected devices pending disposal. 2. If affected devices are identified, contact the recalling firm at support@luminexcorp.com to arrange for replacement and confirm the disposal of any remaining kits in your inventory per the package insert disposal guidelines. 3. Perform a retrospective review of results obtained using the affected SARS-CoV-2 Assay. If you suspect that you have encountered a false negative result with a patient sample, please contact the recalling firm at support@luminexcorp.com. 4. Distribute the notification to any individuals within your organization who need to be made aware of this recall. 5. If affected devices were further distributed, notify your customers of this recall. 6. Please complete and return the Acknowledgement & Receipt Form. 7. For those with complaints, questions, or concerns, please contact Global Technical Support Services at 1-877-785-2323 (U.S. and Canada) or +1-512-381 -4397 (Outside U.S. and Canada). On 12/05/22, additional recall notices were sent to customers notifying them that two additional lots were impacted.
Initiated
2022-09-14
Root cause
Packaging process control
Distribution
Distribution US nationwide, Indonesia, and Hong Kong.
Recall number
Z-0134-2023
Recall status
Open, Classified
CDRH record ID
196236
FDA classified
2022-10-26
Recall event number
90937

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗