Open original FDA record ↗Recall record
- FDA identifier
- Z-0134-2025
- Company in FDA record
- Olympus Corporation of the Americas
- FDA classification date
- Oct 18, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of energy to the patients anatomy.
Source facts
- Action
- Olympus issued Urgent Medical Device Correction Letter on 9/12/24 to Operating Room Director, Risk Management. Letter states reason foe recall, health risk and action to take: 1. Carefully read the content of this notification. 2. Please check all areas of your facility to determine if you have the devices specified above. 3. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification and the Soltive Laser System Instructions for Use. 4. If you have further distributed this product, identify your customers, and forward them this notification. 5. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0452 c. Complete the form as instructed. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@Olympus.com or by phone at (647) 999-3203
- Initiated
- 2024-08-13
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide distribution - US Nationwide and the countries of AU, CA, CL, DE, HK, JP, SG.
- Recall number
- Z-0134-2025
- Recall status
- Open, Classified
- CDRH record ID
- 210101
- FDA classified
- 2024-10-18
- Recall event number
- 95380