Z-0134-2025: Soltive Premium SuperPulsed Laser - Intended for incision, excision, resection, ablation, coagulation, he

Recall record Z-0134-2025 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0134-2025
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Oct 18, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The term Bladder Stone was incorrectly translated in both Spanish and Portuguese to Kidney Stone(Clculo renal) on the systems Graphical User Interface (GUI), a potential to deliver an incorrect or unintended amount of energy to the patients anatomy.

Source facts

Action
Olympus issued Urgent Medical Device Correction Letter on 9/12/24 to Operating Room Director, Risk Management. Letter states reason foe recall, health risk and action to take: 1. Carefully read the content of this notification. 2. Please check all areas of your facility to determine if you have the devices specified above. 3. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification and the Soltive Laser System Instructions for Use. 4. If you have further distributed this product, identify your customers, and forward them this notification. 5. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0452 c. Complete the form as instructed. If you require additional information, please do not hesitate to contact me at Cynthia.Ow@Olympus.com or by phone at (647) 999-3203
Initiated
2024-08-13
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of AU, CA, CL, DE, HK, JP, SG.
Recall number
Z-0134-2025
Recall status
Open, Classified
CDRH record ID
210101
FDA classified
2024-10-18
Recall event number
95380

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗