Skip to content

Public FDA record

Z-0134-2027: JenaValve Trilogy Delivery Catheter, Size 23MM REF: TRILOGY-DC-S

Recall record Z-0134-2027 for Jenavalve Technology Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0134-2027
Company in FDA record
Jenavalve Technology Inc.
FDA classification date
Oct 9, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Delivery Catheter may have sleeve liner separation that results in difficulty or inability to deploy heart valve device.

Source facts

Action
On the dates of August 1, 4, 6, and 19th, JenaValve Technology Inc. issued a Urgent: Medical Device Recall notification via email. JenaValve asked consignees to take the following actions: 1. Review the information contained in this letter and ensure that anyone in your facility who needs to be aware of this notification reads this letter carefully. 2. Examine your inventory immediately to determine if you have Trilogy Delivery Catheters from the affected lots and quarantine the subject products. 3. JenaValve field team members will schedule a visit to retrieve the subject Trilogy Delivery Catheters. If you prefer to return the products yourself, please contact us at 866-588-8287 and we will provide you with an RGA number prior to you shipping them to: JenaValve Technology Inc., 4 Cromwell, Irvine, California 92618, USA; Attn: Quality Department; Reference: FA26001 4. Please complete and sign the Business reply form (BRF) of this letter and email it to customersupport.us@jenavalve.com or give it to your JenaValve field representative. 5. Upon receipt of the completed BRF and returned product, JenaValve will provide replacement products. Incomplete BRFs cannot be processed. 6. If any of the subject Trilogy Delivery Catheters have been transferred to another facility, identify the receiving facility on the Business Reply Form, contact customersupport.us@jenavalve.com, and provide a copy of this letter to the relevant personnel.
Initiated
2026-07-31
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution in the states of AZ, CA, FL, GA, IL, MA, MD, MI, MN, MO, NY, OH, PA, SC, TX, UT, VA, WA and the countries of Germany, Denmark, Great Britain, Italy.
Recall number
Z-0134-2027
Recall status
Open, Classified
CDRH record ID
222212
FDA classified
2026-10-09
Recall event number
99697

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗