Z-0135-2011: ACL TOP (Base) PN 000280000 Bench top, fully automated, random access analyzer designed specifically for

Recall record Z-0135-2011 for Instrumentation Laboratory Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0135-2011
Company in FDA record
Instrumentation Laboratory Co.
FDA classification date
Oct 25, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Software: New data check, "Multiple Threshold Check", was added to identify and fail abnormal (bimodal) clot curves

Source facts

Action
Instrumentaion Laboratory notified the US Distributor/Canada via BCI Service Modification Notice (#10793) on August 13, 2010. The Mandatory software upgrade to Verison V4.3.0 in the United States and Canada will be conducted by BCI service representatives and tracked through the distributor, Beckman Coulter (BCI). International Affiliates/Dealers: Worldwide Regulatory Notices were e-mailed to IL international affiliates and dealers, notifying them of the mandatory software upgrade, providing issue description and a risk summary and the support. The upgrades were to be handled through the local representatives.
Initiated
2010-08-13
Root cause
Software design
Terminated
2012-05-23
Distribution
Worldwide Distribution, USA and Canada
Recall number
Z-0135-2011
Recall status
Terminated
CDRH record ID
94191
FDA classified
2010-10-25
Recall event number
56657

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗