Reason stated in the source

Due to incomplete machining a small peg remnant on the top of the dome that protrudes above the surrounding porous coating causing the cup not to seat properly.

Source facts

Action
On 09/11/2020, the firm sent out an email containing an "URGENT FIELD SAFETY NOTICE" to it Surgical/Sales Agents informing them that a product complaint was received stating that a surgeon had issues with implanting the device and stated that "too tight even after reaming line-to-line" and that the affected device did not seat properly. The Recalling Firm was notifying its agents to immediately return all unused affected devices and pointed out that the risks associated with this issue include the following: -Lack of cup fixation -Device loosening -Fracture -Pain -Delay in Surgery -Revision Surgery. Customers were also to perform the following actions by September 25, 2020: 1. This notice needs to be passed on to all those within your organization or to any organization where the potentially affected product has been transferred. 2. Review your stock for the affected products and lot numbers and to quarantine the items until returned. 3. Contact Customer Service at 1-800-456-8696 to place a replacement order and receive an RMA number. You will be provided with a shipping label to return the affected product. 4. Return all affected devices (quantities listed above) using the RMA number. Report any adverse events related to this issue to Customer Service. Additionally, adverse reactions or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online, by regular mail or by fax. For any further questions, contact Teffany Hutto at teffany.hutto@djoglobal.com and/or contact Customer Service at 1-800-456-8696
Initiated
2020-09-09
Root cause
Employee error
Distribution
U.S. Nationwide distribution including in the states of CA, MN, MS, NY, SC, TN and VA.
Recall number
Z-0135-2021
Recall status
Open, Classified
CDRH record ID
183731
FDA record created
2020-10-20
Recall event number
86450

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗