Skip to content

Public FDA record

Z-0135-2023: LumiraDx SARS-CoV-2 Antigen (Ag) Liquid Quality Control Kit Catalog #: L016080109002 (EUA) The LumiraDx

Recall record Z-0135-2023 for LumiraDx. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0135-2023
Company in FDA record
LumiraDx
FDA classification date
Oct 26, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Assigned vial lot numbers for the incorrect type of product. SARS-CoV-2 & Flu A/B lot numbers were incorrectly assigned to the positive and negative vial lot numbers for SARS-CoV-2 Ag LQCs within batch 2002204017, which should have been assigned SARS-CoV-2 Ag lot numbers resulting in an error message and preventing use of the LQC

Source facts

Action
LumiraDx issued Urgent Product Recall Notice and were provided with an attestation form to document product destruction at their facility via email on 23 September 2022. Letter states reason for recall, health risk and action to take: Immediately stop using the affected lots of LumiraDx SARS-CoV-2 Antigen (Ag) Liquid Quality Controls. " Remove from your stock any remaining affected kits of LumiraDx SARS-CoV-2 Antigen (Ag) Liquid Quality Controls as listed in Table 1. " Ensure any of the affected SARS-CoV-2 Antigen (Ag) Liquid Quality Controls are removed from your inventory are destroyed (not just thrown away), indicate the number of Liquid Quality Control kits that were destroyed, and sign the attestation form at the end of this notice verifying destruction of the affected SARS-CoV-2 Antigen (Ag) Liquid Quality Controls in your inventory. " Report any issues or adverse events with using the LumiraDx SARS-CoV-2 Antigen (Ag) Liquid Quality Controls to LumiraDx immediately or to the FDA via the MedWatch program. " Please forward this notice to additional testing sites if these products were further distributed within your system. " LumiraDx will provide replacement product over the course of the next few days. If you have any questions, contact LumiraDx Customer Services at customerservices.US@lumiradx.com or 1-888-586-4721.
Initiated
2022-09-23
Root cause
Error in labeling
Terminated
2024-12-13
Distribution
US Nationwide distribution.
Recall number
Z-0135-2023
Recall status
Terminated
CDRH record ID
196460
FDA classified
2022-10-26
Recall event number
90998

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗