Open original FDA record ↗Recall record
- FDA identifier
- Z-0136-2014
- Company in FDA record
- Bayer Healthcare Llc
- FDA classification date
- Nov 2, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Test results from these lots may produce a high-bias reading (by up to 20%) that is outside of limits defined for stability and product accuracy when product is stored at room temperature four (4) to five (5) months after release.
Source facts
- Action
- Bayer Healthcare sent an Urgent Medical Device Recall letter dated August 26, 2013. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check their inventory for the affected product and quarantine any affected product. Complete the attached Recall Response Form and fax to Bayer Customer Service. Customers were instructed to use the attached letter to notify affected next level customers of the recall. For questions regarding this recall call 408-524-2255, ext 290.
- Initiated
- 2013-08-26
- Root cause
- Nonconforming Material/Component
- Terminated
- 2014-01-14
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to China, Argentina, Uruguay, Japan.
- Recall number
- Z-0136-2014
- Recall status
- Terminated
- CDRH record ID
- 121482
- FDA classified
- 2013-11-02
- Recall event number
- 66101