Skip to content

Public FDA record

Z-0136-2023: GORE CARDIOFORM ASD Occluder. cardiovascular implant.

Recall record Z-0136-2023 for W L Gore & Associates, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0136-2023
Company in FDA record
W L Gore & Associates, Inc.
FDA classification date
Oct 26, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to manufacturing records (Release Test Results) indicating "Failed".

Source facts

Action
Between 09/21-32/2022, the firm sent an "URGENT MEDICAL DEVICE RECALL/FIELD SAFETY NOTICE" via FedEx 2nd Day (U.S.), Postal Mail (EMEA), and Personal Delivery (Japan), informing customers that W.L. Gore & Associates (Gore) identified devices that were incorrectly released following a documented failure of a subset of quality tests. A failure of the subset of quality tests could potentially result in increased risks during the release of the device from the delivery system for the affected devices. Customer are instructed to: Identify and return any unused devices within the scope of this recall. Complete and sign the enclosed CUSTOMER RESPONSE FORM and return to FieldActionTeam@wlgore.com within 2 weeks of receipt of this notification. Share this letter with others in their institution as appropriate. Transfer this notice to other organization(s) as appropriate. If a listed device has been used, there is no patient follow-up needed and no further actions required other than informing Gore the device was used. Indicate the used device(s) on the CUSTOMER RESPONSE FORM and return to FieldActionTeam@wlgore.com. For questions, contact Gore Customer Service via email - MPDCustomerCare@wlgore.com
Initiated
2022-09-21
Root cause
Employee error
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, CA, CO, DC, FL, GA, IA, ID, IL, ME, MN, NC, NY, JPA, TN, TX, WA, and WI. The countries of Denmark, and Japan.
Recall number
Z-0136-2023
Recall status
Open, Classified
CDRH record ID
196192
FDA classified
2022-10-26
Recall event number
90912

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗