Skip to content

Public FDA record

Z-0136-2025: bk3000, model #2300-11; bk3500, model #2300-41; bk5000, model #2300-61; and bkActive 2300, model #2300-66

Recall record Z-0136-2025 for B-K Medical A/S. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0136-2025
Company in FDA record
B-K Medical A/S
FDA classification date
Oct 18, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The internal power connections at the power inlet of the ultrasound devices can come loose resulting in the potential for excessive heat which could lead to a thermal event or the device becoming inoperable.

Source facts

Action
Notifications were issued on and dated 9/4/2024 informing the consignee of the safety issue and the actions to be taken, which included: Until GE HealthCare implements a permanent correction on the device for this issue, the consignee is to inspect the power cord connection at the power inlet to the ultrasound device. A photograph of the area to be inspected was embedded in the letter. If a cord retaining bracket is not present, the consignee is to ensure the power cord is fully pushed into the power inlet before each use and while the ultrasound device is charging. If a cord retaining bracket is present, no further action is required. The consignee is to ensure all potential staff in the facility are made aware of the safety notification and the recommended actions. An acknowledgment form was enclosed for completion and return.
Initiated
2024-09-04
Root cause
Process control
Distribution
Worldwide - US Nationwide - There was also government/military distribution and the countries of Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Greece, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Republic of Korea, Kuwait, Latvia, Lebanon, Luxembourg, Malaysia, Morocco, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
Recall number
Z-0136-2025
Recall status
Open, Classified
CDRH record ID
210240
FDA classified
2024-10-18
Recall event number
95431

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗