Open original FDA record ↗Recall record
- FDA identifier
- Z-0137-2011
- Company in FDA record
- Instrumentation Laboratory Co.
- FDA classification date
- Oct 25, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Software: New data check, "Multiple Threshold Check", was added to identify and fail abnormal (bimodal) clot curves
Source facts
- Action
- Instrumentaion Laboratory notified the US Distributor/Canada via BCI Service Modification Notice (#10793) on August 13, 2010. The Mandatory software upgrade to Verison V4.3.0 in the United States and Canada will be conducted by BCI service representatives and tracked through the distributor, Beckman Coulter (BCI). International Affiliates/Dealers: Worldwide Regulatory Notices were e-mailed to IL international affiliates and dealers, notifying them of the mandatory software upgrade, providing issue description and a risk summary and the support. The upgrades were to be handled through the local representatives.
- Initiated
- 2010-08-13
- Root cause
- Software design
- Terminated
- 2012-05-23
- Distribution
- Worldwide Distribution, USA and Canada
- Recall number
- Z-0137-2011
- Recall status
- Terminated
- CDRH record ID
- 94193
- FDA classified
- 2010-10-25
- Recall event number
- 56657