Skip to content

Public FDA record

Z-0137-2025: SynchroMed II Infusion Pump. Product Number: 8637-20

Recall record Z-0137-2025 for Medtronic Neuromodulation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0137-2025
Company in FDA record
Medtronic Neuromodulation
FDA classification date
Oct 21, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for error and inability to interrogate pump due to data corruption in the pump memory.

Source facts

Action
On 11-September-2024, a Medtronic representative was sent to the physician s facility where the representative provided the communication and confirmation form to the physician. The physician was asked to sign the confirmation form indicating they read the notification. The representative reset the patient s pump.
Initiated
2024-09-11
Root cause
Device Design
Distribution
US Nationwide distribution in the state of Mississippi.
Recall number
Z-0137-2025
Recall status
Open, Classified
CDRH record ID
210208
FDA classified
2024-10-21
Recall event number
95419

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗