Open original FDA record ↗Recall record
- FDA identifier
- Z-0138-06
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Nov 15, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
If the recommended workflow is not followed and the worksheets are printed after moving the patient table out of isocenter, the offset values will be incorrect. This could result in a biopsy being performed at an incorrect location and presents the possibility of an inconclusive diagnosis. calculation of offset values for the breast biopsy aiming device may be incorrect
Source facts
- Action
- The recalling firm issued a Customer Safety Advisory via certified mail to affected customers per Update Instructions MR034/05/S. The letter re-emphasizes the importance of printing the offset values on the worksheet prior to moving the patient table and in accordance with the User's Manual.
- Initiated
- 2005-10-05
- Root cause
- Other
- Terminated
- 2007-08-07
- Distribution
- The product was shipped to medical facilities in CA, FL, IL, MD, MN, NM, OH, OR, RI, and WI.
- Recall number
- Z-0138-06
- Recall status
- Terminated
- CDRH record ID
- 42746
- FDA classified
- 2005-11-15
- Recall event number
- 33921