Open original FDA record ↗Recall record
- FDA identifier
- Z-0138-2027
- Company in FDA record
- Medtronic Sofamor Danek USA Inc
- FDA classification date
- Oct 9, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Medtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch passed finished product sterility testing Medtronic is retrieving any unused product manufactured from the donor tissue. Donor bone tissue tested positive for Clostridium butyricum and while tissue did undergo gamma radiation, if implanted, there is a possibility of clostridium infection.
Source facts
- Action
- On 8/12/2026, recall notices were mailed or emailed to customers who were asked to do the following: 1) Identify and quarantine any unused impacted product(s). 2) Return affected product(s) per the return instructions in the enclosed Customer Confirmation Form. If product was purchased directly from Medtronic please contact: rs.sdmrarequest@medtronic.com, 3) For patients already implanted with affected products, consider potential Clostridium infection if the patient presents with signs or symptoms of infection. If a patient develops an infection with clostridium at a remote site, evaluate the surgical site for potential infection. 4) Share this Field Safety Notice with all individuals within your organization who need to be aware of this action. If any impacted products have been transferred to another organization, please ensure that a copy of this notice is provided to that organization as well. 5) Complete and return the Customer Confirmation Form via email to neuro.quality@medtronic.com
- Initiated
- 2026-08-12
- Root cause
- Process change control
- Distribution
- US Nationwide distribution in the states of OK, VA, CO, AR, FL, TX, MD, DE, IL, MI, KS, AL, GA.
- Recall number
- Z-0138-2027
- Recall status
- Open, Classified
- CDRH record ID
- 222237
- FDA classified
- 2026-10-09
- Recall event number
- 99701