Periop/Critical Care Manager EHR Software Application displayed one patient's demographic and health information in place of another patient that was in current view.
Source facts
- Action
- Picis Inc. sent a "Safety Notification letter" dated August 12, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The customers were instructed until this service pack is applied, we recommend: * For patients who have been transferred from other facilities within the same VISN, you should discontinue the use of these Picis applications. Manually check the patient's name with that displayed in the Picis patient blue banner to ensure you catch all of these cases. * For all other patients, you may safely continue the use of these Picis applications. For further questions please call (781) 557-3000.
- Initiated
- 2011-08-12
- Root cause
- Software design
- Terminated
- 2012-03-23
- Distribution
- Nationwide Distribution -- AZ, DE, FL, IL, NY, OH, PA, WI and WV.