If digital angiography system is used with a specific catheterization table and the operator directs simultaneous movement of the table and the C-arm using direct memory, then direct memory button is released, the table may continue to move, which the operator can stop by pressing the stop button.
Source facts
Action
On 9/6/24, recall notices were sent to customers who were asked to do the following:
1) Firmware for the system controller requires updating. The work to correct this issue will
be performed by your Authorized Service provider.
2) Schedule time with your Distributer to have your system updated.
3) Complete and return the response form to drodriguez@shimadzu-usa.com
Questions can be directed to the firm's National Technical Support:
Toll Free: (844)487-2767 Available 8:00 to 5:00 PM, Monday - Friday
Initiated
2024-09-06
Root cause
Software Design Change
Distribution
US Nationwide distribution in the states of MA, IL, SC, KY, SC, OH, TN.
Recall number
Z-0139-2025
Recall status
Open, Classified
CDRH record ID
210142
FDA classified
2024-10-22
Recall event number
95406