Z-0139-2027: BD Phoenix Panel NMIC-312. Catalog Number: 449082.

Recall record Z-0139-2027 for Becton Dickinson & Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0139-2027
Company in FDA record
Becton Dickinson & Co.
FDA classification date
Oct 9, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for quality control failures with Klebsiella pneumoniae ATCC BAA-1705 when using the CPO Ambler classification feature.

Source facts

Action
Beckton Dickenson (BD) notified consignees on about 08/13/2026 via mailed and emailed "URGENT: Medical Device Correction" letter. Clinical users were instructed to do the following: 1. Do not report CPO Ambler classification results generated from affected BD Phoenix CPO detect 6-well panels. 2. Do not release Ambler classification information from BD Phoenix panels on patient reports, laboratory information system reports, electronic health records, or other clinical reporting outputs. 3. CPO positive results, other resistance marker results, and any antimicrobial AST MIC and/or SIR result from the BD Phoenix panels are unaffected and may still be reported according to routine laboratory reporting practice. 4. Confirmatory testing to verify appropriate Ambler Classification should be determined and performed in accordance with established laboratory procedures. 5. There are no recommendations regarding results review or sample retesting. Customers were instructed to ensure the contents of the notification are read and understood, share and post the customer letter within each facility network, forward to any customers to whom product may have been distributed to. Additionally, customers were instructed to complete and return the provided Customer Response Form. BD is investigating the root cause and will implement appropriate corrective actions to rectify this issue as soon as possible. **UPDATE 09/02/2026: Firm expanded recall and sent updated notification to include an additional product code. Instructions remained the same.
Initiated
2026-08-13
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-0139-2027
Recall status
Open, Classified
CDRH record ID
222297
FDA classified
2026-10-09
Recall event number
99725

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗