Open original FDA record ↗Recall record
- FDA identifier
- Z-0140-2010
- Company in FDA record
- Kentucky Packaging Service LP
- FDA classification date
- Nov 10, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
THERE WERE INCONSISTENCIES NOTICED ON THE INSULATION OVERLAP. THE ELECTRODE HAS A LARGE DIAMETER SHAFT THAT TRANSITIONS TO A SMALLER DIAMETER SHAFT. THE 1ST LAYER OF INSULATION WHICH COVERS THE SMALL DIAMETER SHAFT OVERLAPS WITH THE 2ND LAYER OF INSULATION THAT COVERS THE LARGER DIAMETER SHAFT. ON SOME ELECTRODES IT APPEARS THAT THE OVERLAP IS TOO SHORT TO COVER THE TRANSITION WHICH MAKES THIS
Source facts
- Action
- Olsen Medical issued an "Urgent: Device Recall" letter to Consignees beginning September 24, 2009 . Customers were instructed to inventory their stock and for all remaining product immediately discontinue use and return. Consignees were also instructed to contact their customers and notify them of the recall. For further information, contact Olsen Medical at 1-502-772-4272.
- Initiated
- 2009-09-24
- Root cause
- Process control
- Terminated
- 2010-10-05
- Distribution
- Nationwide Distribution -- AZ, FL, IA, MD, MO, NM, OR, NV, TN, TX and VA.
- Recall number
- Z-0140-2010
- Recall status
- Terminated
- CDRH record ID
- 85595
- FDA classified
- 2009-11-10
- Recall event number
- 53482