Philips Medical Systems North America Co. Phillips
FDA classification date
Oct 25, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
Reliability of the Stand-Trolley cable may result in loss of images on the monitor during an examination. This intermittent problem can lead to System lock-up, where no radiation is possible and images are not visible on the monitor
Source facts
Action
On 10/04/2008, Philips began sending the URGENT-Device Correction BV Pulsera R2.2 and Endura R2.2 letters to their customers.
On 09/03/2008, the firm issued the Field Change Order (FCO) 71800024 and the firm's representative will replace the stand-trolley cable free of charge.
Customers should contact Philips Call Center at 1-800-722-9377, #5, #2, and reference "FCO 71800024" with any questions.
Initiated
2008-09-05
Root cause
Component design/selection
Terminated
2010-12-07
Distribution
Nationwide distribution.
Recall number
Z-0140-2011
Recall status
Terminated
CDRH record ID
73916
FDA classified
2010-10-25
Recall event number
49605