Z-0140-2014: Siemens RAD Fluoro Uro system (Luminos dRF and Uroskop Omnia with software version VD10A/G Image intensi

Recall record Z-0140-2014 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0140-2014
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Nov 27, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When using systems operating with software versions VD10A/G during a RAD examination, a malfunction can result in radiation continuing longer than required resulting in an overexposed image that is not of diagnostic quality, prompting repetition of the examination.

Source facts

Action
Siemens sent an Field Safety Notice dated September 3, 2013, to all affected customers.The letter identified the product the problem and the action to be taken by the customer. Action to be taken: 1. It is not possible for the user to prevent this malfunction. The exposure time with the use of automatic exposure control is limited to 500ms, which is the maximum duration of radiation in this mode. The exposure can be interrupted by the user at anytime by releasing the exposure button. 2. There are two possibilities to solve this issue: a.) Switch off the system if the message "wait for FLC" in status bar does not disappear approximately. 30s after the OGP selection. The system gets ready after switching the system on and reselecting the Torno OGP. b.) Press Shift+F10. The pop up message has to be confirmed and you have to wait approximately. 40s until! the "i" icon in the status bar stops blinking. After that a RAD OGP has to be selected and then the Torno OGP. Thereafter, the system gets ready. 3. It is not possible for the user to prevent this malfunction The Center for Devices and Radiological Health (CDRH) understands that your Corrective Action Plan (CAP) will involve 59 units and consist of the following elements: 1. You have notified affected customers with Field Safety Notice XP032/13/S. 2. You have prepared a field modification of the Luminos dRF and Uroskop Omnia systems with software version VD10A/G. 3. Distribution of this field modification will began in September 2013 as Update Instruction XP015/13/S at no cost to the user. For further Customer Support call (610) 219-6300.
Initiated
2013-09-03
Root cause
Software design
Terminated
2015-03-09
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada.
Recall number
Z-0140-2014
Recall status
Terminated
CDRH record ID
122033
FDA classified
2013-11-27
Recall event number
66314

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗