Z-0141-2027: Brand Name: IOMAX Product Name: IOMAX Cortical Module Model/Catalog Number: 190296-200 UDI-DI code: 00840

Recall record Z-0141-2027 for Cadwell Industries Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0141-2027
Company in FDA record
Cadwell Industries Inc
FDA classification date
Oct 9, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Stimulator device contains a defective printed circuit board assembly (PCBA) that may cause the device to unintentionally deliver electrical stimulation to the patient, or allow an unintended electrical current to reach the patient.

Source facts

Action
On 08/18/2026, an "URGENT: MEDICAL DEVICE RECALL" Letter was emailed informing customers that, a manufacturing defect was identified in a limited number of IOMAX Cortical Modules. Following a recent change in printed circuit board assembly (PCBA) vendor, the defect was discovered during routine production processes on a limited number of PCBA s that were used in IOMAX Cortical Modules. The identified risk is specific to the Cortical Module and Cascade Surgical Studio (CSS). However, because Cortical Modules can be used with Zenith, Arc, and CSS, a module currently in use with Zenith or Arc could later be connected to CSS. Cadwell is therefore recalling all potentially affected Cortical Modules regardless of current host system. If the defective PCBA is present, it could potential cause unintentionally deliver electrical stimulation to the patient, or allow a small amount of unintended electrical current to reach the patient. Customers are instructed to return all identified devices to Cadwell Industries, Inc. Cadwell Industries, Inc. is voluntarily recalling and replacing or repairing all affected devices from the field.
Initiated
2026-08-18
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the state of TN and the countries of Spain, Vietnam, Malaysia, Saudi Arabia, Thailand, Israel, Mexico, Turkey, Oman, Colombia, Brazil, Belgium, Italy, South Africa, and Switzerland.
Recall number
Z-0141-2027
Recall status
Open, Classified
CDRH record ID
222448
FDA classified
2026-10-09
Recall event number
99774

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗