Open original FDA record ↗Recall record
- FDA identifier
- Z-0141-2027
- Company in FDA record
- Cadwell Industries Inc
- FDA classification date
- Oct 9, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Stimulator device contains a defective printed circuit board assembly (PCBA) that may cause the device to unintentionally deliver electrical stimulation to the patient, or allow an unintended electrical current to reach the patient.
Source facts
- Action
- On 08/18/2026, an "URGENT: MEDICAL DEVICE RECALL" Letter was emailed informing customers that, a manufacturing defect was identified in a limited number of IOMAX Cortical Modules. Following a recent change in printed circuit board assembly (PCBA) vendor, the defect was discovered during routine production processes on a limited number of PCBA s that were used in IOMAX Cortical Modules. The identified risk is specific to the Cortical Module and Cascade Surgical Studio (CSS). However, because Cortical Modules can be used with Zenith, Arc, and CSS, a module currently in use with Zenith or Arc could later be connected to CSS. Cadwell is therefore recalling all potentially affected Cortical Modules regardless of current host system. If the defective PCBA is present, it could potential cause unintentionally deliver electrical stimulation to the patient, or allow a small amount of unintended electrical current to reach the patient. Customers are instructed to return all identified devices to Cadwell Industries, Inc. Cadwell Industries, Inc. is voluntarily recalling and replacing or repairing all affected devices from the field.
- Initiated
- 2026-08-18
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide - US Nationwide distribution in the state of TN and the countries of Spain, Vietnam, Malaysia, Saudi Arabia, Thailand, Israel, Mexico, Turkey, Oman, Colombia, Brazil, Belgium, Italy, South Africa, and Switzerland.
- Recall number
- Z-0141-2027
- Recall status
- Open, Classified
- CDRH record ID
- 222448
- FDA classified
- 2026-10-09
- Recall event number
- 99774