Open original FDA record ↗Recall record
- FDA identifier
- Z-0142-2014
- Company in FDA record
- Sorin Group Deutschland GmbH
- FDA classification date
- Nov 4, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Sorin Group is recalling certain models of S5 Perfusion Systems due to malfunction of cardioplegia control.
Source facts
- Action
- Sorin Group sent an Urgent Field Safety Notice dated September 24, 2013, to all affected customers. The letter informed users of the potential for failing of automatic cardioplegia delivery function and notified them that they would be contacted by the Sorin Service Group to arrange an appointment to update the firmware on site. Customers were asked to complete the Response form to confirm they had received, read and understood the Field Notice. For questions regarding this recall call 303-467-6527.
- Initiated
- 2013-09-24
- Root cause
- Software Manufacturing/Software Deployment
- Terminated
- 2014-07-10
- Distribution
- Worldwide Distribution - USA including WA, NY, VA, TN, SC, TX, CA, CO, MA, WI, MD, AL, ND, and Puerto Rico and Internationally to Italy, Finland, Germany, Sweden, and Great Britain.
- Recall number
- Z-0142-2014
- Recall status
- Terminated
- CDRH record ID
- 122435
- FDA classified
- 2013-11-04
- Recall event number
- 66464