Z-0142-2024: Centricity Universal Viewer Workflow Manager, Image processing radiological system

Recall record Z-0142-2024 for GE Healthcare. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0142-2024
Company in FDA record
GE Healthcare
FDA classification date
Oct 23, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When using Universal Viewer or Centricity Universal Viewer with Workflow Manager (WFM), if a third-party reporting application is used to launch exams, it could display information on a different patient than WFM and the Viewer.

Source facts

Action
GE Healthcare notified consignees on about 09/08/2023 via letter. Consignees were instructed that until the software update is available the device can still be utilized through an updated browser setting or through discontinuing the use of bi-directional reporting. A GE Healthcare representative will contact each consignee to arrange for the correction. Consignees were instructed to complete and return the provided acknowledgement response.
Initiated
2023-09-08
Root cause
Software design
Distribution
Nationwide distribution to AR, AZ, CA, CO, DE, FL, GA, IL, IN, ME, MI, MN, MO, NE, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX. International distribution to Austria, Canada, France, Germany, Hong Kong, Israel, Italy, Philippines, Singapore, Thailand, Turkey, Vietnam.
Recall number
Z-0142-2024
Recall status
Open, Classified
CDRH record ID
203341
FDA classified
2023-10-23
Recall event number
93053

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗