Open original FDA record ↗Recall record
- FDA identifier
- Z-0143-2024
- Company in FDA record
- GE Healthcare
- FDA classification date
- Oct 23, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When using Universal Viewer or Centricity Universal Viewer with Workflow Manager (WFM), if a third-party reporting application is used to launch exams, it could display information on a different patient than WFM and the Viewer.
Source facts
- Action
- GE Healthcare notified consignees on about 09/08/2023 via letter. Consignees were instructed that until the software update is available the device can still be utilized through an updated browser setting or through discontinuing the use of bi-directional reporting. A GE Healthcare representative will contact each consignee to arrange for the correction. Consignees were instructed to complete and return the provided acknowledgement response.
- Initiated
- 2023-09-08
- Root cause
- Software design
- Distribution
- Nationwide distribution to AR, AZ, CA, CO, DE, FL, GA, IL, IN, ME, MI, MN, MO, NE, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX. International distribution to Austria, Canada, France, Germany, Hong Kong, Israel, Italy, Philippines, Singapore, Thailand, Turkey, Vietnam.
- Recall number
- Z-0143-2024
- Recall status
- Open, Classified
- CDRH record ID
- 203342
- FDA classified
- 2023-10-23
- Recall event number
- 93053